Endodontic instruments fracture in root canal: Integrative review

Pâmela de Medeiros Dantas1; Renata Correia Sotero Dália Torres1; Robeci Alves Macedo Filho1; Sandra Aparecida Marinho2*

1 Instituto de Odontologia das Américas (IOA)/ Instituto de Odontologia da Paraíba (IOP), Brazil.

2 Curso de Odontologia, Universidade Estadual da Paraíba (UEPB), Campus VIII, Brazil.

*Corresponding Author: Sandra Aparecida Marinho
State University of Paraiba (UEPB), Cel Pedro Targino Street sn, 58233-000, Araruna, PB, Brazil.
Email: [email protected]

Received : Sep 06, 2021

Accepted : Oct 14, 2021

Published : Oct 21, 2021

Archived : www.jcimcr.org

Copyright : © Marinho SA (2021).

Abstract

Introduction: Endodontic instruments fracture is related to several factors from internal dental anatomy and type, to operator inability and inexperience.

Objective: To review case reports related to fracture of endodontic instruments inside the root canal.

Methodology: A literature review was carried out using papers available in the PubMed and Lilacs databases. Only case reports were included, with no restriction on language and publication year

Results: It was analyzed eight papers with 11 instrument fracture cases, with the highest prevalence (45.5%) of rotary instruments fracture, occurring mainly in middle-cervical and middle-apical root canals thirds. In 90.9% of cases, techniques used for removal showed successful results and in only one case (9.1%), the technique was unsuccessful and the instrument was kept inside the root canal.

Final considerations: Endodontic instruments fracture is a complication can directly affect prognosis and endodontic treatment success. Several techniques and devices can be used to remove fractured fragments inside root canals, however, there is no specific protocol for their removal.

Keywords: accidents; endodontics; fracture; dental instruments; prognosis.

Citation: Dantas PDM, Torres RCSD, Filho RAM, Marinho SA. Endodontic instruments fracture in root canal: Integrative review. J Clin Images Med Case Rep. 2021; 2(5): 1374.

Introduction

Instruments fracture during root canal treatment is a complication during endodontic treatment [1], which can occur due instrument torsion or its cyclic fatigue and/or operator inability. In rotary instruments, torsional fracture occurs when tip or any other instrument part locks into the canal while the rest continues to rotate until it fractures. Flexural fadigue, on the other hand, manifests itself when repeated compression and tension movements occur in a root canal at its flexion maximum point [2].

Stainless steel hand files have been widely exclusively used to canal preparation for a long time, however, these instruments can cause steps, deviations and canal perforations due their high rigidity and low elasticity. To minimize complications and overcome anatomical challenges of root canals, rotary Nickel-Titanium (NiTi) instruments were introduced in the market. Rotary NiTi files have high flexibility and elasticity, greater mechanical strength and less tendency to rectify the canals, when compared to stainless steel files [3]. Despite advantages of NiTi instruments, especially in curved canals, is relatively high, compromising treatment success. The prevalence of stainless steel instruments fracture during the canal exploration was 1.39% and using NiTi rotary instruments was 2.77% [4].

Faced a fracture instrument, the chances of successful removal must be evaluated for potential complications. Clinical conduct and treatment prognosis will depend on fragment type and size, fracture site and its accessibility, internal anatomy and pulp condition and also canal preparation at the time of instrument fracture, which reflects microbial control. Three approaches can be used to solve intracanal instrument fracture: remove or bypass the instrument, or prepare and fill the root canal over the fractured fragment [5,6].

Operating microscope is a valuable resource, as it allows a direct visualization of root canal and fractured instrument [7], thus presenting better results in instrument removal [8].

The aim of this study was to carry out an integrative review of case reports covering endodontic instrument fractures inside the root canal.

Methodology

This is an integrative literature review of case reports, referring to factors related to endodontic instrument fracture inside root canal. Only observational case report studies were included, with no restriction on language and publication year.

Search was carried out in November 2020, in electronic databases PUBMED (http://www.pubmed.gov) and LILACS (www. bireme.br), using the following descriptors: [(“endodon- tics”) AND (“fractured” instrument”) AND (“Case Report”)]. After reading papers titles/abstracts, those which met eligibility criteria were saved. After reading the full text, the references that confirmed the eligibility criteria fulfillment were included in this review.

Data contained information about study country, imaging examination, affected tooth, fractured instrument location and type, removal technique, use of magnification, follow-up and case outcome were collected.

Results

Results are shown in Table 1.

Cases reported in Brazil, India, Australia and Mexico were observed in eight papers, involving 11 teeth affected by instru- ment fractures, which eight (72.7%) were molars and three (27.3%) were anterior teeth. Periapical radiography was performed in 100% of the cases, and only one study associated cone beam computed tomography with this exam.

Table 1: Case Reports of instruments fractured inside the canal included in the review.

 

AUTHOR,
COUNTRY

 

TEETH

 

IMAGINOLOGICAL EXAM

 

INSTRUMENT
LOCATION

 

TYPE OF INSTRUMENT

 

REMOVAL
TECHNIQUE

 

USE OF
MAGNIFICATION

 

FOLLOW UP

 

 OUTCOME

 

Ward (2003), Austrália

 

16

 

Periapical Radiography

 

Curvature of Apical Third (MB conduct)

 

2.5mm fracture of Profile 25.04 file

 

Ultrasound technique variation describe by Ruddle

 

Microscope (8x).

 

12 months Assintomatic tooth, with no periapical  radiolucence

 

Successful technique, allowing preparation, desinfection and filling of conducts

 

 

Thirumalai et al. (2008), Índia

 

 

Case 1: 16

Case 2: 36

Case 3: 22

 

 

Periapical Radiography

 

Case 1: Coronal  and  Middle Thirds (DB conduct)

Case 2: Coronal  and Middle Thirds (DL conduct)

Case 3: Middle  and Apical Thirds

Case 1: Fracture of 5mm Hedstroem #25 file

Case 2: Fracture of hand file stainless steel #K25

Case 3: Fracture of hand file NiTi #35

 

 

Masserann Technique

 

 

NO

 

 

NO

Successful technique. A straight-line acess to the fragment facilited centering the trephan. Locking mechanism of the extractor provides retention in gripping and dislodging a fragment which is tightly wedged in the canal.

 

Rahimi; Parashos (2009), Austrália

 

46

 

Periapical Radiography

 

Curvature of Apical Third (DB conduct)

 

3mm fracture of  Race rotary instrument

 

chloroform- dipped gutta percha cone

 

NO

 

NO

 

Conservative and safe technique for removing
loose fractured instruments in  hard-to reach areas

 

Choksi et al (2013), Índia

 

12

 

Periapical Radiography

 

Coronal and  Middle thirds

 

Not informed

 

Masserann Technique

 

NO

 

NO

 

Endodontic treatment ended with no complications

 

Brito Jr et al. (2014), Brasil

 

 

37

 

Periapical Radiography Cone Beam Computed Tomography (Follow up)

 

Middle and Apical thirds (MB conduct)

 

10mm fracture of S1 ProTaper Rotary instrument

Gates Glidden drills, Endo z bur and bypassing the instrument without sucess. Instrument left in the canal

 

NO

 

3 years. Clinical and radiographic findings with absence of periapical radiolucency

 

Successful Treatment.
The presence of fractured instrument did not impact the prognosis during the follow up period

 

Chinna et al. (2015), Índia

 

 

36

 

 

Periapical Radiograph

 

Coronal and  Middle Thirds (DB conduct)

 

10mm Fracture of   HeroShaper instrument

 

Ultrasonic Tip  ET25

 

NO

 

 

6 months. Radiographic findings with no alterations.

 

Conservative and safe Technique  for removing
fractured instruments

 

Cruz et al. (2015), México

 

36

 

Periapical Radiograph

 

 

Apical third (Distal conduct), crossing the foramen

 

Radiopaque area suggesting fractured instrument (#15 or #20 file)

Modified Gates Glidden drills,  ultrasonic tips #4 (Pro-Ultra, Dentsply), and IRS, that was removed with the instrument

 

Microscope (6x)

 

2 years. Asymptomatic
Patient. Disappearance of small radiolucency in distal conduct

 

Operative microscope + ultrasound+ IRS system by staging platform technique aollowed the removal of instrument beyond the foramen

 

Brito Jr et al. (2015), Brasil

 

 

 

Case 1: 21

Case 2: 37

 

Periapical Radiograph

Case 1: Middle and Apical thirds

Case 2: Middle and Apical thirds.
(MB conduct)

Case 1: Bur mettalic fragments

Case 2: F1 ProTaper Rotary instrument

Case 1: Fragments removal with costumized extractor

Case 2: Modified injection needle and steel wire

 

NO

 

NO

 

Cases 1 e 2: Successful removals

Regarding fractured fragments location, four (36.4%) instruments fractured in the coronal/middle third; four (36.4%) in the middle/apical third and three (27.2%) instruments in the apical third of the root canals.

Of 11 fractured instruments, five (45.5%) were rotary files (one Profile 25.04 File, one Race Rotary File, one HeroShaper Rotary File and two S1 Protaper Files); two (18.2%) were stainless steel hand files (Hedstroem File #25 and Kerr File #25); one (9.1%) was a manual NiTi file (#35) and one (9.1%) was related to metallic drill fragments. In two (18.2%) cases, the type of fractured instrument was not identified.

To remove the fragments, the Masserann technique was used in four (36.4%) cases, ultrasonic tips were used in three (27.3%) cases (operating microscope was used in only two cases - 18.18%). Alternative methods were used in the other cases (27.3%), such as softening gutta percha cone with chloroform, a modified extractor and a modified anesthetic needle with metallic steel wire. The techniques used showed good results in 10 (90.9%) cases with the fragments being removed. The technique was unsuccessful in only one case (9.1%) and the fragment reimaned inside the canal. Clinical and radiographic findings revealed absence of signs and symp- toms in followed up cases.

Discussion

Technological innovations in NiTi files have enabled a better root canal system biomechanical preparation [9], as they allowed an increase in taper, thus causing root canal widening with greater efficiency in removing debris and increased flow of irrigating solution. However, NiTi files associated with a rotating system increase their canal adhesion and cause difficulties in fragment removing in case of fracture [10].

During endodontic treatment, stainless steel instruments often fracture due excessive amounts of torque and NiTi in- struments fracture due combined action of tension, twist and cyclic loading. However, stainless steel files are easier to re- move, as they do not fragment during the process [3]. When compared to continuous rotation, reciprocating rotary systems provide greater resistance to fracture, as they increase cyclic fa- tigue of the NiTi instrument [11]. In rotary NiTi instruments, tor- sional failure, which may be caused by using too much apical force during instrumentation, occurred more frequently than flexural fatigue, which may result from use in curved canals [2].

Endodontic Instrument fractures may be related to several factors, such as internal anatomy and incorrect access to root canal; instrument confection material and manufacturing process and inadequate use. According to Parashos et al. [5], variables related to the operator and canal anatomy have more influence on fractures than the instrument itself. A root canal with a high degree of curvature, added to an incorrect technique increase the possibility of instrument fracture. In addition, insufficient access without adequate dentin removal from coronal cavity walls makes it difficult to properly view and locate the canals, resulting in inadequate instrument orientation, which negatively affects cyclic fatigue and strength, generating stress concentration and fracture predisposition [12].

Clinical management and treatment prognosis of an instrument fractured will depend on pulp condition (vital or nonvital), signs and symptoms, internal anatomy (curvature degree), and level of canal cleaning and shaping; presence of periapical lesion; type, size, visualization and accessibility to the fragment, in addition to chemical-mechanical debridement before fracture, reflecting microbial control [13,14]. Fractured instrument position, whether in apical, middle or cervical thirds of the root can also interfere in the prognosis [15].

According to each case specificity, instrument can be removed or kept inside the canal, or can be performed bypass [6]. Bypass allows root canal preparation with instrument incorporation to the filling material [12]. C hances of successful removal must be evaluated carefully considering potential complications. Tooth prognosis can be seriously compromised due excessive dentin wear to recover the instrument, predisposing root fractures and irreversible tooth loss [12,13].

Instrument fractures do not always lead to an unfavourable prognosis. Instrument removal from apical third of curved canals should not be routinely attempted [16]. Fractured fragment may not cause treatment failure when pulp is vital without periapical involvement. When the instrument is located in a difficult access place, where its remotion can cause severe root structures damage, bypass technique is indicated [17]. After three years of follow-up using this technique, Brito Jr. et al. [17] observed an absence of signs and symptoms. On the other hand, if the instrument is not removed or it has passed the canal foramen in an infected and necrotic pulp, the prognosis will be less favorable [9,18].

Technologies integration such as ultrasound magnification (optical microscope and magnifying glass) and computed tomography are valuable resources during the removal planning in instrument fracture cases. They allow better instrument visualization and selective root canal shaping. That association presents better results in instruments removal when compared to manual methods [8]. Cone beam computed tomography allows more accurate dental morphology assessment and endodontic complications diagnosis, highlighting fractured instruments location [19]. Chinna et al. [20] demonstrated the ultrasound is a conservative and predictable technique for removing fractured instruments. However Brito Jr et al. [17] were not successful with ultrasonic tips to remove instruments in root apical third. Despite the success rate, using ultrasonic removal also poses risks, with some accidents such as fragment transposition to another location, perforation and fracture of the root. The more apical the instrument location, the greater perforation risk [21].

Several alternative techniques have been applied for different clinical situations. Among them, Masserann kit, adhesives (cyanoacrylate), Canal-Finder system, Instrument Removal System (IRS) and use of surgical hypodermic needle [12]. Masserann kit has been used for over 50 years, with highest success rate for anterior teeth, which have thick and straigh roots. It has an end-cutting trephan burs of increasing size which are rotated anticlockwise to create space around fragment coronal end by cutting the surrounding root canal dentin. Also it contains a hollow tube that is subsequently positioned around the fragment [22,23]. However, it has limited application in thin and curved roots teeth or instrument fracture in the apical region, as the use of relatively large and rigid trephans leads to considerable root dentin removal, generating tooth weakening or perforation risk [24].

Brito Jr et al. [12] developed a modified injection needle technique to reduce dentin wear. It was a conservative and low-cost method for removing intracanal instruments. Other alternative method was proposed by Cruz et al. [25], who used a dental operating microscope, ultrasound and the IRS system combined, which allowed endodontic instrument removal located beyond the foramen and enabled a conventional disinfection protocol. Ward [26] used a staging platform and ultrasonic tips with direct visualisation using an operating microscope, to remove a fractured rotary NiTi instrument from the apical third of a curved root canal. The definitive management should be based on a thorough knowledge of success rates of each treatment option balanced against potential removal risks or file retention [6].

Final considerations

Factors related to endodontic instruments fracture are: internal dental anatomy, mechanical fragility of instruments due constant use and operator inability. Fragment location, type and size; instrumentation phase and periapical condition directly influence protocol choice and treatment prognosis. Several techniques and therapeutic modalities are available for fractured instruments recovery from root canal, however, without having satisfactory results with a single protocol. Associated use of microscope and ultrasonic systems can be a promising option. Further studies should be carried out to establish more conservative and safer protocols for the fractured instruments removal.

References

  1. Lin C et al. A statistical model for predicting the retrieval rate of separated instruments and clinical decision-making. J Dent Sci. 2015; 10: 423-30.
  2. Sattapan B, et al. Defects in rotary nickel-titanium files after clinical use. J Endod. 2000; 26: 161-5.
  3. Gundogar M, Ozyurek T. Cyclic fatigue resistance of OneShape, HyFlex EDM, WaveOne Gold, and Reciproc Blue Nickel-titanium Instruments. J Endod. 2017; 43: 1192-6.
  4. Tavares WLF. et al. [Índice de fratura de instrumentos manuais de aço inoxidável e rotatórios de NiTi em clínica de pós-graduação em Endodontia.] Arq Odontol. 2015; 51.
  5. Parashos P et al. Factors influencing defects of rotary nickel-titanium endodontic instruments after clinical use. J Endod. 2004; 30: 722-5.
  6. McGuigan MB et al. Clinical decision-making after endodontic instrument fracture. Br Dent J. 2013; 214: 395-400.
  7. Van Der Sluis LWM et al. Passive ultrasonic irrigation of the root canal: a review of the literature. Int Endod J. 2007; 40: 415-26.
  8. Gencoglu N, Helvacioglu D. Comparison of the different techniques to remove fractured endodontic instruments from root canal systems. Eur J Dent. 2009; 3: 90-5.
  9. Berman L, Hargreaves KM. Cohen’s Pathways of the pulp expert consult. Mosby. 2010; 1008p.
  10. Prateek J et al. Management options of intracanal separeted instruments: A review. J Pharm Sci Innov. 2013; 2: 17-21.
  11. Barbosa-Ribeiro M et al.Canal transportation and centering ability of curved root canals prepared using rotary and reciprocating systems. Braz J Oral Sci. 2015; 14: 214-8.
  12. Brito-Júnior M et al. Alternative techniques to remove fractured istrument fragments from the apical third of root canals: Report of two Cases. Braz Dent J. 2015; 26: 79-85.
  13. Choksi D et al. Management of an intracanal separated instrument: A case report. Iran Endod J. 2013; 8: 205-7.
  14. Rover G et al. Influence of access cavity design on root canal detection, instrumentation efficacy, and fracture resistance assessed in maxillary molars. J Endod. 2017; 43: 1657-62.
  15. Suter B et al. Probability of removing factured instrument from root canals. Int Endod. 2005; 38: 112-23.
  16. Rahimi M, Parashos P. A novel technique for the removal of fractured instruments in the apical third of curved root canals. Int Endod J. 2009; 42: 264-70.
  17. Brito-Júnior M et al. Obturation over an S1 ProTaper instrument fragment in a mandibular molar with three years of follow-up. Braz Dent J. 2014; 25: 571-4.
  18. Panitvisai P et al. Impact of a retained instrument on treatment outcome: a systematic review and meta-analysis. J. Endod. 2010; 36: 775-80.
  19. Brito RCA et al. Detection of fractured endodontic instruments in root canals: Comparison between different digital radiography systems and Cone-beam Computed Tomography. J Endod. 2017; 43: 544-9.
  20. Chinna H et al. Ultrasonics: A novel approach for retrieval of separated instruments. J Clin Diag Res. 2015; 9: 18.
  21. Shahabinejad H et al. Success of ultrasonic technique in removing fractured rotary Nickel-Titanium endodontic instruments from root canals and its effect on the required force for root fracture. J Endod, 2013; 39: 824-8.
  22. Okiji T. Modified usage of the Masserann kit for removing intracanal broken instruments. J Endod. 2003; 29: 466-7.
  23. Thirumalai AK, et al. Retrieval of a separated instrument using Masserann technique. J Conserv Dent. 2008; 11: 42-5.
  24. Pai AR, Kamath MP, Basnet P. Retrieval of a separated file using Masserann technique: A case report. Kathmandu Univ Med J. 2006; 4: 238-42.
  25. Cruz A et al. Removal of an Instrument fractured by ultrasound and the instrument removal system under visual magnification. J Contemp Dent Pract. 2015; 16: 238-42.
  26. Ward JR. The use of an ultrasonic technique to remove a fractured rotary Nickel-Titanium instrument from the apical third of a curved root canal. Austr Endod J. 2003; 29: 25-30.